Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 19
Page 19 of 68
Medical Solutions
Alert 0
Tab. 3 Service Alert Table
Code Reply
keys
Message
-6 OK #1:Alert
#2:System info error
#3:A/D board info is updated.
[Cause]
The A/D board information is different from the former informa-
tion. This problem occurs when the version of the A/D board has
been updated or when the connected units have been changed.
If this error occurs when no such changes have been made, it is
attributed to another problem. In this case, a service person
needs to be called in.
[Operation after the reply key is touched] The A/D board infor-
mation is renewed.
-8 OK #1:Alert
#2:System self-check error
#3:Call a service person and let him/her know the error code.
[Cause]
An error has been found by a self-check.
[Operation after the reply key is pressed]
Although imaging is possible, image may be degraded. There-
fore, a service person needs to be called in.