20 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 20 of 68
Medical Solutions
-116 OK #1:Alert
#2:Sensor unit:Self-check error
#3:Detected an error.
#4:ER(n) (xxxx) n: Sensor index (1 to 4) xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed] Nothing
-5002 OK #1:Alert
#2:OPU-CCR RPC Error
#3:OPU-CCR RPC Error
[Cause]
For some reason, the CCR software has shut down.
[Operation after the reply key is pressed]
Code Reply
keys
Message