Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Backup and restore 59
Page 59 of 68
Medical Solutions
Restore of a drive partition 0
Restore is required only if the imaging system is not working properly.
Restore c: + d: is performed when problems occur in the Win XP operating system or
imaging system application software.
After restoring a drive partition, the current organ programs and current customer adapta-
tions must be reinstalled using the last imaging system backup CD of the configuration
(refer to the "Backup/Restore of the Configuration" chapter).
Preliminary step:
Procedure
Switch off the MOBILETT XP Digital.
Open the top cover of the MOBILETT XP Digital, lay it sideways on a table or the like
and access the PC interior. Do not disconnect any cables.
Open the PC so that it is possible to use the PC keyboard.
Switch on the MOBILETT XP Digital and press “F2” on the keyboard during boot.
¹ Access the BIOS mode of the computer.
Enter the BIOS password “99999”.
In the menu, verify that the first boot device is set to DVD ROM; change it if necessary.
Save changes and exit the menu.
Insert the CD MOBILETT XP Digital system DVD into the CD RW drive and boot from
the DVD.
¹ The following menu opens:
- 1 - Back up the complete hard drive (c: + d:)
- 2 - Restore the complete hard drive (c: + d:)
- 3 - Back up the complete hard drive (c:
- 4 - Restore the complete hard drive (c:
- 5 - Use GHOST interactive
- E - Exit
¹ Drive c: contains the operating system, drive d: contains the application soft-
ware.
Enter “2” and confirm with “Enter”; wait for a message to appear.
Following the message window,
¹ A DOS window opens at the end of the restore operation.This concludes the
system restore of the image system.
Enter “E” to exit.
The configuration data must now be restored to reinstall all organ program and custom-
er adaptations. Refer to the "Backup/Restore of the Configuration" chapter.
Switch off the MOBILETT XP Digital.
- The MOBILETT XP Digital system DVD is required.
- The last MOBILETT XP Digital backup CD "Install#" is required.