Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Backup and restore 61
Page 61 of 68
Medical Solutions
To save data to CD, Windows XP "Write data to CD" from Windows Explorer has to be
used.
Select “SERVICE TOOL > Explorer” in the menu.
Label an empty backup CD with the date and "Backup of the Configuration" and num-
ber it.
Insert the labeled "Backup of the Configuration" CD into the CD drive.
In Explorer, copy folder "D:\_Siemens_\Configuration" onto drive e:.
Right-click on drive e: folder and select "Write these files to CD"
Following the messages,
¹ All configuration data from _Siemens_ saved to CD.
Wait until the burning process is ready.
Close Explorer.
Exit the service tool window via “Exit”.
Exit administrator setup via “OK".
Switch off the MOBILETT XP Digital
Save all backup CDs to a secure place within the MOBILETT XP Digital
Reinstall the PC and the top cover.
Restore of the configuration 0
The most current complete "Backup of the Configuration" CD has to be used to restore
the configuration data. Use Internet Explorer to copy the configuration data previously
copied to the backup CD to the D:\TEMP folder. All configuration data and organ pro-
grams are restored with the imaging system “SERVICE TOOL/restore configuration to
ccr-folder".
Preparation:
Procedure
Switch off the MOBILETT XP Digital.
Open the top cover of the MOBILETT XP Digital, lay it sideways on a table or the like
and access the PC interior. Do not disconnect any cables.
Open the PC so that it is possible to use the keyboard of the PC.
Switch on the Mobilett XP Digital.
From the user interface of the MOBILETT XP Digital:
Select “SYSTEM > CONFIGURATION”.
Select “ADMINISTRATOR SETUP > SERVICE TOOL”.
Click on the password box
¹ - Enter the administrator password and confirm with “OK” (refer to the password
list).
- The last MOBILETT XP Digital "Backup of the Configuration" CD is required.