62 Backup and restore
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 62 of 68
Medical Solutions
Select “SERVICE TOOL > Explorer” in the menu.
Start via “START”.
¹ Enter the Siemens service password and confirm with “OK” (refer to the pass-
word list).
Insert the most recent labeled "Backup of the CONFIGURATION" CD into the CD drive.
Use Windows Explorer to copy the "e:\Configuration" folder to drive d:\TEMP.
Close Windows Explorer.
Restore of configuration files
Select “SYSTEM > CONFIGURATION”.
Select “ADMINISTRATOR SETUP > SERVICE TOOL”.
Click on the password box
¹ - Enter the administrator password and confirm with “OK” (refer to the password
list).
Select “SERVICE TOOL > restore configuration to ccr-folder” in the menu.
Start via “START”.
¹ Enter the Siemens service password and confirm with “OK” (refer to the pass-
word list).
¹ All required files are copied to the D:\CCR directory.
Close the window via “OK".
Exit the “SERVICE TOOL” window via “Exit”.
Exit “ADMINISTRATOR SETUP” via “OK".
Switch off the MOBILETT XP Digital
Reinstall the PC and the top cover.
Restart the system.