62 Backup and restore
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 62 of 68
Medical Solutions
• Select “SERVICE TOOL > Explorer” in the menu.
• Start via “START”.
¹ Enter the Siemens service password and confirm with “OK” (refer to the pass-
word list).
• Insert the most recent labeled "Backup of the CONFIGURATION" CD into the CD drive.
• Use Windows Explorer to copy the "e:\Configuration" folder to drive d:\TEMP.
• Close Windows Explorer.
Restore of configuration files
• Select “SYSTEM > CONFIGURATION”.
• Select “ADMINISTRATOR SETUP > SERVICE TOOL”.
• Click on the password box
¹ - Enter the administrator password and confirm with “OK” (refer to the password
list).
• Select “SERVICE TOOL > restore configuration to ccr-folder” in the menu.
• Start via “START”.
¹ Enter the Siemens service password and confirm with “OK” (refer to the pass-
word list).
¹ All required files are copied to the D:\CCR directory.
• Close the window via “OK".
• Exit the “SERVICE TOOL” window via “Exit”.
• Exit “ADMINISTRATOR SETUP” via “OK".
• Switch off the MOBILETT XP Digital
• Reinstall the PC and the top cover.
• Restart the system.