Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 39
Page 39 of 68
Medical Solutions
-5102 Grid Error: Remove grid for this
exposure.
Touch OK and remove the grid from
the detector unit.
-5202 Grid ID Error: Replace with the
recommended grid (name of
grid). Installed grid: (name of
grid).
Touch OK and check the grid in the
detector unit.
-5203 Grid ID Error: This is a non-grid
mode. Remove the grid.
Touch OK and remove the grid from
the detector unit.
-5206 Grid ID Error: This is a fixed-grid
mode. Attach the grid.
Touch OK and install the grid on the
detector unit.
Error Code Error Message Remedy