38 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 38 of 68
Medical Solutions
-513 Media Control Error: Operation
has been canceled because the
external storage or directory is
busy.
• Touch OK and try again after a
while.
-514 External Storage Error: Exposed
images cannot be stored
because capacity of folder parti-
tion is smaller than size of
exposed images. Use a new
disk, erase the folder contents,
or increase the folder size and
retry.
• Touch OK and set the Max Size of
Disk.
-517 Internal Storage Error: Pro-
cessed image cannot be stored
because capacity of internal
storage memory is insufficient.
• Touch OK and shut down the sys-
tem .
• If these errors occur, contact a Sie-
mens representative or distributor.
-518 Internal Storage Error: DICOM
image cannot be stored because
capacity of internal storage
memory is insufficient.
-519 External Storage Error: Capac-
ity of external storage memory is
insufficient. Retry after making
space on the disk.
• Touch OK. Replace the external
disk and try again.
-520 Internal Storage Error: Exposed
image cannot be stored because
capacity of internal storage
memory is insufficient.
• Touch OK and shut down the sys-
tem.
• If these errors occur, contact a Sie-
mens representative or distributor.
-521 Queue File Open Error: Queue
file not found.
-599 File Open Error: File not found.
-601 Resend Error: File may have
been automatically deleted.
• Touch OK.
-5100 Detector Unit: Detect Error:
Check the detector power unit or
cable connection. If system
recovery is desired, shut down
system.
• Touch OK and shut down the sys-
tem.
• Check cable connections.
• If the cause cannot be found, con-
tact a Siemens representative or
distributor.
-5101 Detector Unit: Detect Error:
Check the detector power unit or
cable connection. If system
recovery is desired, shut down
system.
Error Code Error Message Remedy