Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 17
Page 17 of 68
Medical Solutions
-105 OK/ -97 #1:Fatal error
#2:CXDCAP.INI Open error
#3:Install CXDCAP.INI file.
[Cause]
The cxdcap.ini file has not been installed.
[Operation after the reply key is pressed]
The system enters a fatal condition. All the sensor units will
therefore be disconnected. A service person needs to be called
in. However, you can transfer the image and use the study list.
-97
Cxdcap.ini file not found.
-522 OK #1:Fatal error
#2:File Open Error:filename
#3:File is not found.
[Cause]
For some reason, XXXX.BIN and XXXX.DCM files do not exist.
Or, they cannot be read.
[Operation after the reply key is pressed]
Shutdown starts. Although this problem can be solved using the
service tool, users must not use the service tool on their own.
Take necessary actions in accordance with the service person"s
instruction.
Code Reply
keys
Message