Error messages of the imaging system 27
Page 27 of 68
Medical Solutions
-302 OK #1:Error
#2:Calibration error
#3:Too much dose. Or collimator is used.
[Cause]
Excessive dose of radiation was used in exposure of image for
white correction.
[Operation after the reply key is pressed]
Perform imaging again.
-303 OK #1:Error
#2:Calibration error
#3:Too little dose. Or collimator is used.
[Cause]
Insufficient radiation dose was used in exposure of image for
white correction.
[Operation after the reply key is pressed]
Perform imaging again.
-304 OK #1:Error
#2:Calibration error
#3:Invalid data. Or collimator is used.
[Cause]
Too much difference in dose acquired by the panels.
[Operation after the reply key is pressed]
Perform imaging again.