52 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 52 of 68
Medical Solutions
Tab. 10 Customer Attention Table
Error Code Error Message Remedy
-506 DICOM Response Warning:
Although transfer finished prop-
erly, receiving device reported a
warning in DICOM response.
Check receiving device.
• Touch OK and check the settings of
the output destination.
-508 DICOM Printer Status Warning:
Although transfer finished nor-
mally, receiving device reported
a WARNING in DICOM printer
status. Check receiving device.
• Touch OK and replace the external
disk before the capacity becomes
insufficient. ** is the value you have
set.
-511 External Storage Attention:
Remaining space in external
storage is less than **% .