54 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 54 of 68
Medical Solutions
After a command
request was sent
to the server, an
error was returned
in response.
0x10 CXDI description: DICOM response error
(CCRTRANS_ERR_RESP)
The log output in this case is shown below.
-------------------------------------------------------------------
-(Not determined
After a command
request was sent
to the server, a
warning was
returned in
response.
0x18 CXDI description: DICOM response warning
(CCRTRANS_WRN_RESP)
The transfer process was successful, but a warn-
ing was returned from the server.
The log output in this case is shown below.
-------------------------------------------------------------------
-(Not determined)
The printer status
has returned an
error.
0x20 CXDI description: DICOM printer status error
(CCRTRANS_ERR_PRN_STATUS)
The log output in this case is shown below.
-------------------------------------------------------------------
-(Not determined)
The printer status
has returned a
warning.
0x28 CXDI description: DICOM printer status warning
(CCRTRANS_WRN_PRN_STATUS)
The transfer process was successful, but a warn-
ing was returned as the printer status.
The log output in this case is shown below.
-------------------------------------------------------------------
-(Not determined)
Other errors Values other
than those
above
CXDI description: DICOM communication
error(CCRTRANS_ERR_DICOM_TRANSE)
The log output in this case depends on the specific
error. A typical example is shown below.
-------------------------------------------------------------------
-18-140933[d2]ERR : c0082 SRV Send (DATA
SET) failed in SRV_SendDataSet
18-140933[d2]ERR : 190082 SRV Request failed
in SRV_NCreateRequest
18-140933[d2]ERR : 70012 NULL_key passed to
routineDUL_ReleaseAssociation
Sample error Return value CXDI description/ log output