EasyManua.ls Logo

Siemens MOBILETT XP Digital - Mechanical Function Test; Cable Winch; Support Arm Movements; Single Tank and Collimator Movements

Siemens MOBILETT XP Digital
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR8-230.814.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Unpacking and visual inspection 29
Page 29 of 70
Medical Solutions
Mechanical function test 0
Cable winch 0
P Pull out the cable until it is completely unwound.
Hold the cable with one hand near the duct to slow it down as it is being wound. Pull up
the brake of the cable winch and wind up the cable completely.
This device is equipped with a standard safety plug. Attach a power plug that is compliant
with local standards if necessary.
Support arm movements 0
P Unlock the support arm. Then check its movements, including several stops. If the DAP
chamber option is included in the delivery, it must be attached to the collimator. The arm
should be easy to move through its full range using one hand and should stop in every
position without any additional movement (up or down).
Fig. 11: Positioning of support arm
Single tank and collimator movements 0
P Turn the collimator with the handle to the end positions ±90°.
Rotate the tube to all of its commonly used working and park positions.

Table of Contents

Other manuals for Siemens MOBILETT XP Digital

Related product manuals