• The patient leakage current must be measured at the flat detector.
• The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
P
• Enter the value determined as the first measured value in the patient leakage current
report (chapter 11 of these instructions).
• This value may not exceed the permissible leakage current values according to DIN
VDE 0751-1/ 2001-10, table F.1, line "patient leakage current" of 0.01 mA for direct cur-
rent and 0.1 mA for alternating current.
P
• The measuring procedure and the measuring device used (designation and serial num-
ber) must also be documented.
Labels 0
Country-specific labeling
P
• The "Warning!" label "This X-ray unit may be dangerous...” is supplied in a number of
languages. Select the correct language and attach the label according to the MOBI-
LETT XP Digital operator manual.
• The "Danger!” label “Explosion hazard...” is supplied in a number of languages. Select
the correct language and attach the label according to the MOBILETT XP Digital opera-
tor manual.
• Affix the "Do not open doors to rooms with the unit" label as shown in (Fig. 27 / p. 55).
• For Germany only:
Select the 1.5 m, 2.5 m or 3.5 m "control area" label as required by the customer and
affix it to the bottom of the arm as shown in (Fig. 28 / p. 56).