Siemens AG SPR8-230.814.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Start-up 41
Page 41 of 70
Medical Solutions
Options 0
DAP 0
P Measuring device for dose area product (Dose Area Product).
The DAP measuring device is used to measure the kerma area product (kerma = kinetic
energy released in matter) during an X-ray examination. It is primarily used to record the
radiation dose a patient is exposed to.
The DAP is calibrated at the factory:
• Press the test button on the DAP display, a value between 80-120 µGym
2
(DAP resolu-
tion 0.1 µGym
2
) or in the case of a high-resolution chamber, 8-12 µGym
2
(DAP resolu-
tion 0.01 µGym
2
) must be subsequently displayed.
If the display is incorrect, the measuring device must be calibrated. See the
"Replacement of Parts" instructions, SPR8-230.841.30...).
X
• Check the function of the DAP display by releasing an exposure. The DAP display
shows a measurement value depending on the object in the beam path.
Remote control 0
P • Adhere the remote control holder to the front right or left side (16/15/Fig. 3 / p. 12).
• Switch on the MOBILETT XP Digital.
X
• Check the "collimator light" and "exposure release" functions of the remote control.