Siemens AG SPR8-230.814.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
DHHS 43
Page 43 of 70
Medical Solutions
Test of kV accuracy 0
Requirement:
The measured kV values must comply with the limit values specified in the tables. One of
two measuring methods can be used:
1. kV meter procedure:
- kV meter based on the filter comparison technique (e.g. PTW-Nomex). Use the cor-
rection factor given in the kV meter operating instructions for non-invasive kV mea-
surements.
2. Oscilloscope method:
- Storage oscilloscope with a measuring accuracy of ± 2.5%
kV meter method: 0
NOTE
The inherent filtration (Al equivalent) of MOBILETT XP Digital is
3.1mm Al (single tank plus collimator).
An installed DAP measurement chamber increases the inherent fil-
tration by 0.4 mm AL.
Switch the MOBILETT XP Digital on.
Maintain or switch to manual operation of the MOBILETT XP.
Prepare the kV meter for the measurement according to the operating instructions.
Place the measuring detector on a suitable surface and adjust the single tank to the de-
tector using the light localizer (Fig. 18 / p. 44).
Set the source-to-image distance to the value specified in the kV meter operating in-
structions.