44 DHHS
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 44 of 70
Medical Solutions
Fig. 18: kV measurement
X • Set the following exposure parameters and release an exposure after each setting:
* The measuring inaccuracy of the measuring instrument used must be subtracted from
the specified limit values.
P Record the measured values.
Oscilloscope method: 0
"KVS" measuring point for the nominal value and “KV” measuring point for the actual
value. The measuring ratio is 30 kV/V.
• Remove the top cover and place on a table or similar surface. All cables remain con-
nected. Measuring points on CPU D916 (see troubleshooting instructions).
• Connect the oscilloscope to the measuring points of the CPU D916 TP KV, KVS, and
GND (also see the troubleshooting instructions). Oscilloscope setting: Channel 1 = TP
KV 1V/div, channel 2 TP KVS 2V/div, trigger channel 2, trigger stage 2.5 V, 2V/div, 50
ms/div
kV meter
Selection
Exposure parameters Limit values*
DC voltage
52 kV 50 mAs 49.0 - 55.0 kV
81 kV 20 mAs 77.0 - 85.0 kV
133 kV 12.5 mAs 126.4 - 139.6 kV