44 DHHS
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 44 of 70
Medical Solutions
Fig. 18: kV measurement
X Set the following exposure parameters and release an exposure after each setting:
* The measuring inaccuracy of the measuring instrument used must be subtracted from
the specified limit values.
P Record the measured values.
Oscilloscope method: 0
"KVS" measuring point for the nominal value and “KV” measuring point for the actual
value. The measuring ratio is 30 kV/V.
Remove the top cover and place on a table or similar surface. All cables remain con-
nected. Measuring points on CPU D916 (see troubleshooting instructions).
Connect the oscilloscope to the measuring points of the CPU D916 TP KV, KVS, and
GND (also see the troubleshooting instructions). Oscilloscope setting: Channel 1 = TP
KV 1V/div, channel 2 TP KVS 2V/div, trigger channel 2, trigger stage 2.5 V, 2V/div, 50
ms/div
kV meter
Selection
Exposure parameters Limit values*
DC voltage
52 kV 50 mAs 49.0 - 55.0 kV
81 kV 20 mAs 77.0 - 85.0 kV
133 kV 12.5 mAs 126.4 - 139.6 kV