MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 34 of 70
Medical Solutions
Fig. 15: Holder
Collimator function 0
P Completely open the collimator.
Check the functioning of the light keys on the collimator.
Check the manual collimations with the light on.
The light switches off after approx. 20 sec.
Test the collimator light key on the control panel (1/Fig. 13 / p. 32).