56 Remaining work
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 56 of 70
Medical Solutions
Fig. 28: "Control area" label
Reports 0
Separate the completed 'Start-up report' (chapter 9 of these instructions) from these in-
structions and file it in the system binder, Register 9, "Certificates."
Separate the completed protective conductor resistance report, leakage current report,
and patient leakage current report (chapter 11 of these instructions) from these instruc-
tions and file them in the system binder, Register 9, "Certificates."
Complete the "Installation Protocol" included with the unit and send it (by fax) to the ad-
dress listed in the protocol.
Separate the completed DHHS report (chapter 10 of these instructions) from
these instructions and file it in the system binder, Register 9, "Certificates."