• Determination for the following exposure parameters (product of tube current x expo-
sure time):
X
* The measuring inaccuracy of the measuring instrument must be subtracted from the
specified limit values.
P Calculate and record the mAs values.
mAs meter method: 0
Use of the mAs jumper on PCB D907 directly at the single tank.
• Remove cover from single tank and connect mAs meter to D907 (1/Fig. 20 / p. 47).
• Follow the same procedure as described in “Oscilloscope method”.
For more information on the measuring technique, refer to the user's manual for the
mAs meter.
Fig. 20: mAs jumper D907
Control unit setting
Allowable mAs value*
kV mAs
40 5 4,7 - 5,3
81 2 1,9 - 2,1
133 10 9,5 - 10,5