Test of mAs accuracy 0
Requirement:
The measured mAs values must comply with the limit values specified in the tables. One
of two measuring methods can be used:
1. mAs meter method:
- mAs meter (e.g. MAS meter 8160400 with a measuring accuracy of 1% ± 1 digit)
2. Oscilloscope method:
- Storage oscilloscope with a measuring accuracy of ± 2.5%
Oscilloscope method: 0
Close the collimator.
Remove the top cover and place on a table or similar surface. All cables remain con-
nected.
Clamp the oscilloscope to the “JR” and “GND” measuring points on CPU D916.
Switch the MOBILETT XP Digital on and select manual operation.
After switching the unit on, wait for approx. 5 minutes to let it stabilize.
Example: Calculation of the mAs value
Fig. 19: mAs calculation
The mAs value can be calculated with the help of the oscilloscope display.
Surface “A” corresponds to the mAs value.
mAs = tube current x exposure time
mAs = 340 mA x 0.006 s = 2.0 mAs