EasyManua.ls Logo

Siemens MOBILETT XP Digital - Backup Procedure; Backup of Hard Disks

Siemens MOBILETT XP Digital
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR8-230.814.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Backup 51
Page 51 of 70
Medical Solutions
8-
7Backup
Backup procedure 0
Included with delivery of the MOBILETT XP Digital is a backup of the entire C: driveand D:
drive on the system DVD (factory settings).
P Following start-up of the MOBILETT XP Digital at the customer site and completion of all
adjustment and configuration steps, a second backup operation of the hard drives to CD
is performed.
A "Backup of Configuration" is also created. The "Backup of Configuration" includes all
configuration settings, organ programs, log files, and images from the external storage.
This ensures that the customer configuration can be restored or that the imaging system
can be reinstalled with the customer configuration.
Backup of hard disks 0
Prerequisite
The Mobilett XP Digital has been completely installed and configured according to cus-
tomer specifications. All checks have been successfully completed.
The MOBILETT XP Digital System DVD is available.
At least 2 blank writable CDs are available (2 blank CDs are included with the delivery).
Delete all test exposures in “local Temp Storage” and in “External Storage” prior to the
backup.
Procedure
The MOBILETT XP Digital is switched off.
Open the top cover of the MOBILETT XP Digital and place on a table or similar surface.
All cables remain connected (access to CXDI PC).
Open the CXDI PC (laptop) to access the keyboard and display. Do NOT switch on the
CXDI PC if it is disconnected from the MOBILETT XP Digital. Otherwise, the display
setting can switch to the default setting without touch screen display on the laptop.
Label the blank CDs used for the backup with the date and image and number them se-
quentially (min. 2 CDRs required)
Switch the MOBILETT XP Digital "ON" and place the system DVD in the DVD/RW drive
of the CXDI PC (boot from the system DVD)
NOTE
If necessary, set the boot device to DVD-ROM in the laptop's BIOS.
Access to BIOS: while system is booting, press the F2 key and en-
ter the administrator password "99999".
¹ The display shows the following menu:
- 1 - Backup the complete hard drive (c: + d:)
- 2 - Restore the complete hard drive (c: + d:)
- 3 - Restore factory settings (c: + d:)
- 4 - Back up the complete hard drive (c:)