54 Remaining work
MOBILETT XP Digital SPR8-230.814.30.03.02 Siemens AG
11.05 CS SD 24
Page 54 of 70
Medical Solutions
9-
8Remaining work
Final activities 0
Protective conductor test 0
Observe the protective conductor resistance test information in these instructions.
• Perform the protective conductor test with all covers closed in accordance with
ARTD-002.731.17... The protective conductor resistance must not exceed 0.2 Ohms.
P
• The values determined must be recorded as initial measurements in the protective con-
ductor resistance report (chapter 11 of these instructions), and the measuring points
must be given.
P
• The measuring procedure and the measuring device used (designation and serial num-
ber) must also be documented.
Leakage current measurement 0
Observe the leakage current measurement information in these instructions.
• The leakage current must be measured with all covers closed, in accordance with
ARTD-002.731.17... The limit value of 2.5 mA must not be exceeded.
WARNING
Electrical voltage!
Non-compliance can lead to severe injury and even death.
¹ No housing parts of the system may be touched during
direct measurement of the leakage current .
¹ Third-person access to the system must be prevented.
• The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
P
• Enter the highest value as the first measured value in the leakage current report (chap-
ter 11 of these instructions).
• This value must not exceed the permissible leakage current values according to DIN
VDE 0751-1/2001-10, Table F.1, line "leakage current for devices according to re-
marks 1 and 3", of 2.5 mA.
P
• Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a mea-
surement at the nominal value of the line voltage. This must also be documented.
P
• The measuring procedure (differential measurement or direct measurement) and the
measuring instrument used (designation and serial number) must also be documented.
Patient leakage current measurement 0
Observe the patient leakage current measurement information in these instructions.
• The patient leakage current must be measured with all covers closed, in accordance
with ARTD-002.731.17...