Siemens AG SPR8-230.814.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Backup 53
Page 53 of 70
Medical Solutions
Writing backup data to CD
After backing up all relevant data to the D:\_Siemens_ folder using the imaging system
service tool, the data in this folder has to be written to CD using Windows XP Explorer.
Label the blank CD for the backup with the date and "Backup of Configuration" and
number it.
Select “Explorer” from the SERVICE TOOL menu.
Start via "START"
¹ Enter the service password (see password list)
Insert the labeled CD into then CD RW drive of the CXDI PC.
Copy the “D:\_Siemens_” folder to drive E:\ (CD RW drive).
Right-click in the right pane of the E:\ drive Explorer window and select "Write these
files to CD".
Follow the prompts.
¹ The data is saved to CD
Wait until the write operation is completed.
Close Explorer.
Click “Exit” to exit the “SERVICE TOOL”.
Click “OK” to exit “ADMINISTRATOR SETUP”.
Switch the MOBILETT XP Digital OFF.
Close the CXDI PC, secure the CXDI PC transport fastener, and attach the top cover