3.12 Patient leakage current measurement
SIE Patient leakage current measurement
Observe the patient leakage current measurement information in these instructions.
Perform the patient leakage current measurement with all covers closed, according to
ARTD-002.731.17....
Measure the patient leakage current at the flat detector.
The limit values of 0.01 mA for direct current and 0.1 mA for alternating current may not
be exceeded.
The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
Document the measuring procedure and the measuring device used (designation and
serial number).
The value determined must be entered in the patient leakage current report and evalu-
ated.
NOTE
File the patient leakage current report in the "Certificates" register
no. 9 in the system binder.
NOTE
A new report must be created if the patient leakage current mea-
surements are not documented.
A new patient leakage current report can be found in the startup in-
structions SPR8-230.814.30.03. Separate this report from these in-
structions, fill it out, and file it in the "Certificates" register no. 9 in
the system binder. If values are newly determined, they must be re-
corded as first measured values. The evaluation is omitted in this
case.