Inspection and maintenance 59
Page 59 of 74
Medical Solutions
3.6 Options
3.6.1 DAP measuring system
Measuring device for dose area product (Dose Area Product).
The DAP measuring device is used to measure the kerma area product (kerma = kinetic
energy released in matter) during an X-ray examination. It is primarily used to record the
radiation dose a patient has been exposed to.
The DAP is calibrated at the factory.
SIE Function of the DAP measuring system
• System is switched on.
• Press the Test button on the DAP display.
¹ The subsequently displayed value must be between 80-120 µGym
2
(DAP reso-
lution range 0.1 µGym) or 8-12µGym (DAP resolution range 0.01µGym) for a
high-resolution chamber.
¹ If the display is incorrect, the measuring device must be calibrated. See the
"Replacement of parts" instructions (SPR8-230.841...).
X
• Verify the function of the DAP measuring system by releasing an exposure.
¹ A measured value is displayed in the DAP display as a function of the object in
the beam path.
3.6.2 Remote control
PMP Battery replacement for remote control (yearly)
• Replace the battery (9V alkaline).
SIE Remote control function
• Check the function of the remote control.
• Switch on the light localizer of the collimator.
X
• Initiate an exposure.