62 Inspection and maintenance
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 62 of 74
Medical Solutions
Connect the oscilloscope to the measuring points of the (CPU) D916 TP ”KV” and
“KVS” (also see the troubleshooting instructions). Oscilloscope setting: Channel 1 = TP
KV 1V/div, channel 2 TP KVC 2V/div, trigger channel 2, trigger stage 2.5 V, 2V/div, 50
ms/div (oscilloscope setting: Channel 1 = TP KV 1V/div, channel 2 TP KVC 2V/div, trig-
ger channel 2, trigger stage 2.5 V, 2V/div, 50 ms/div)
¹ Oscilloscope setting: Channel 1 = TP "KV" 1V/div, channel 2 TP "KVS" 2V/div,
trigger channel 2, trigger stage 2.5 V, 50 ms/div
Follow the same procedure as described in “kV meter method”.
3.7.2 mAs accuracy
Requirement:
The measured mAs values must be within the limit values, see tables.
One of two measuring methods can be used:
1.) mAs meter method:
2.) Oscilloscope method:
Oscilloscope method:
Open the top cover of the MOBILETT XP Digital and place it sideways on a table or the
like. Do not disconnect any cables.
Connect the oscilloscope to the "JR" and "GND" measuring points on CPU D916.
Switch on the MOBILETT XP Digital.
Select the analog operating mode.
¹ After switching the unit on, wait approx. 5 minutes for it to stabilize.
X
Release exposures according to the table.
- mAs meter (e.g., MAS meter 8160400 with a measuring accuracy of 1 % ± 1
decimal place)
- Storage oscilloscope with a measuring accuracy of ± 2.5%
Parameter setting Allowable mAs value
kV mAs
40 5 4.7 - 5.3
81 2 1.9 - 2.1
133 10 9.5 - 10.5