18 General maintenance information
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 18 of 74
Medical Solutions
2.3 Material replacement
Replace damaged or missing screws.
NOTE
Only replace damaged or missing screws with steel screws.
All Allen screws must have a tensile strength of 8.8.
Materials needed for periodic replacement can be requested from the responsible local
Uptime Service Center.
Material Material no. Interval
Collimator lamp 08392016 12 months
Battery pack (left) BK1/BK2 08392024 24 months
Battery pack (right) BK3/BK4 08392032 24 months
Cable winch complete 06508746 48 months
Option
Remote control batteries 1x9V alkaline 12 months