Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
General maintenance information 17
Page 17 of 74
Medical Solutions
2.2 Required tools, test equipment and aids
• Standard service tool kit
• Measuring device for leakage current and patient leak-
age current, e.g.:
51 38 727 Y0766
• Ground wire tester or e.g.: 51 38 727 Y0766
• 2.1 mm precision copper filter (or alternatively, copper
filter set 4406 120 RV090)
90 00 598 RV090
• Dose meter, e.g.: PTW DIADOS 97 17 612 Y0388
• Densitometer, e.g., DensiX-LE 52003 49 51 286 Y0388
• Digital multimeter, e.g.: Fluke 8060 A 97 02 101 Y4290
• Centering cross 96 60 051 RE999
• mAs meter (e.g.: 81 60 400) or 2-channel storage oscil-
loscope with ± -2.5% accuracy
• kV meter (works with the filter comparison method) or
2-channel storage oscilloscope with +/- 2.5% accuracy
• 2-channel storage oscilloscope with 2.5% accuracy,
e.g.: Fluke Scope Meter 199 CM
73 92 074
• Spring balance for 350 N
• Torque wrench for 8-40 Nm, e.g.: 99 00 846 RE999
• Calibration tool for the "DAP measurement system" op-
tion
65 84 978
• Adjuster for tensioning the drive belt 65 64 301
• Rope (min. length 4 m; min. tensile strength 1000 N)
• Loctite 242
• Viscogen oil
• 2 pieces of wood, approx. 50 x 75 x 500 mm
• Lux meter (USA only)