Disassembly 2 - 7
Siemens AG SPR2-240.841.01 Page 7 of 12 POWERMOBIL
Medical Solutions Rev. 01 05.03 CS PS 24
Removing the C-arm and replacing the C-arm cable 2
• Block the cable module. For this purpose, move the screw (1/Fig. 12) from the thread
hole (2/Fig. 12) to the thread hole (1/Fig. 12) and tighten it.
• Remove the holders of the covers of the cable modules on both sides (1/Fig. 13).
• Have 2 persons turn the C-arm out of the bearing, as shown in Fig. 14. For this purpose,
try to take the weight off the C-arm to avoid excess pressure to the C-arm from the last
rollers.
• Set the C-arm down.
• The cables in the interior of the C-arm are included in the scope of supply of the cable
module. Replace them one-to-one. When pulling out the old cables draw in a pull wire.
Fig. 12
Fig. 13 Fig. 14