POWERMOBIL SPR2-240.841.01 Page 10 of 10 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
3 - 10 Installation
• When running cables along the frame for the PC boards D1 and D30 (Fig. 22), make
sure that they are not crushed when inserting and tightening the frame.
• Fix the run cables using cable ties.
• Refit the plate (3/Fig. 21).
C-arm cover 3
• Secure the front C-arm cover to the image intensifier side of the C-arm (Fig. 23).
When inserting cables make sure that they are not crushed or torn off.
Covers of the horizontal carriage 3
• Refit the cover plate (2/Fig. 24).
• Reattach the cable in the cable clamps. Make sure that the shielding makes good ground
contact.
• Screw the protective ground wire to the cover of the horizontal carriage.
• Reconnect the plugs of the brake button and of the Emergency Stop switch.
• Lift the cover above the horizontal carriage and fasten it with screws (1/Fig. 24).
Fig. 23 Fig. 24