POWERMOBIL SPR2-240.841.01 Page 8 of 12 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
2 - 8 Disassembly
Cables within the energy chain 2
• Remove the plate (4/Fig. 15).
• Remove the fabric tape from both ends of the energy chain.
• Open the energy chain fully (1/Fig.15).
• Remove the Plexiglas cover (3/Fig. 15).
• Release the frame of the PC boards D1 and D30 on the opposite side of the basic unit,
and pull it out to the side (Fig. 16).
• Remove the ferrite core from the M2 X1 cable (5/Fig. 15).
• Trace the cables inside the energy chain, and loosen the corresponding cable ties,
cable clamps and plug/screw connections in the basic unit.
• Pull any cables back towards the energy chain to facilitate threading the cables in and
out.
Fig. 15 Fig. 16