Siemens AG SPR2-240.841.01 Page 1 of 12 POWERMOBIL
Medical Solutions Rev. 01 05.03 CS PS 24
Disassembly 2
Preparations 2
• Position the basic unit in the room so that it can be accessed from all sides.
• Lock the foot brake.
• Push the squared timber underneath the basic unit as indicated by the arrow in Fig. 1 to
prevent the basic unit from tilting back after removing the POWERPHOS and the X-ray
image intensifier.
• Switch the POWERMOBIL on, and move the lifting column to its approximate vertical
center position.
• Switch the POWERMOBIL off again, and unplug the monitor cable X10 plug from the ba-
sic unit.
• Unplug the monitor cart's power plug to completely disconnect the monitor cart from the
power supply.
Fig. 1 Fig. 2