POWERMOBIL SPR2-240.841.01 Page 8 of 10 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
3 - 8 Installation
I.I. mini voltage supply and X-ray image intensifier 3
• Connect the Lemosa plug connector in the C-arm. Insert the plug connector into the
C-arm. Run the cable in the C-arm so that it is positioned at the side underneath the met-
al cover (1/Fig. 19).
• Insert the control and high-voltage sections of the I.I. mini voltage supply in the C-arm.
• Plug in the connection cable (ribbon cable) between control section and high-voltage
section.
• Attach the X-ray image intensifier in the (suspended) service position.
• Fasten the ground wire connections to the image intensifier.
• Connect the high-voltage cable as well as the cables E1, E2 and E3.
• Connect the black cathode cable of the X-ray image intensifier to the CON1.8 plug.
• Connect the CON1 plug to the control section.
• Attach the camera plug M4 X1 to the mounting bracket.
• Connect the plug M30 X1.
• Fasten all cables using cable ties.
• Lift the X-ray image intensifier, and carefully thread in the cables. Guide the high-voltage
cable, the cable E3 and the protective ground wire through the bracket (1/Fig. 20).
Insert the ribbon cable between control section and HV section of the I.I. mini voltage
supply, at the side between the metal cover (1/Fig. 19) and the C-arm.
• Fasten the image intensifier.
Fig. 19 Fig. 20
To avoid blocking camera rotation after fastening the X-ray image
intensifier make sure that all cables are run carefully during
installation.
NOTE