POWERMOBIL SPR2-240.841.01 Page 4 of 12 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
2 - 4 Disassembly
POWERPHOS 2
If applicable, dismantle the DIAMENTOR dose measuring chamber and loosen the con-
nector. Also remove the insulating disk underneath the DIAMENTOR dose measuring
chamber.
Remove the cover above the collimator. The collimator remains attached to the
POWERPHOS.
Remove the fan component from the POWERPHOS.
Remove the cable ties of the fan cable.
Note down the connection order of the fan assembly supply cable, and loosen the ca-
bles (terminals and colors of the cables, see 1/Fig. 6).
Also loosen the cable for the fan integrated in the C-arm. Also note down the connection
order (1/Fig. 6).
Fig. 6 Fig. 7
Fig. 8