POWERMOBIL SPR2-240.841.01 Page 4 of 12 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
2 - 4 Disassembly
POWERPHOS 2
• If applicable, dismantle the DIAMENTOR dose measuring chamber and loosen the con-
nector. Also remove the insulating disk underneath the DIAMENTOR dose measuring
chamber.
• Remove the cover above the collimator. The collimator remains attached to the
POWERPHOS.
• Remove the fan component from the POWERPHOS.
• Remove the cable ties of the fan cable.
• Note down the connection order of the fan assembly supply cable, and loosen the ca-
bles (terminals and colors of the cables, see 1/Fig. 6).
• Also loosen the cable for the fan integrated in the C-arm. Also note down the connection
order (1/Fig. 6).
Fig. 6 Fig. 7
Fig. 8