• Run the cables in the POWERPHOS C-arm segment so that the cables are not crushed
when inserting the POWERPHOS.
• Run the loose contacts to the plug Z66 X1, and reinsert them (Fig. 17).
Tie the cables using cable ties.
• Reinsert the POWERPHOS in the C-arm segment. Make sure that the cables are not
crushed.
• Screw the two inverter connectors back on. Tighten the nuts with a torque of 4.8 Nm.
• Fasten the protective ground wires.
• Connect all plugs to the POWERPHOS.
• Connect the collimator plug.
• Attach the sterile cover above the collimator.
• If applicable, assemble the DIAMENTOR dose measuring chamber with insulating disk
and attach it.
• Attach the two cable clamps (1/Fig. 18). Make sure that the cable shields makes good
ground contact.
• Fit the fan assembly, and connect it observing the connection order as previously noted.
• Tie or attach all cables using cable ties.
• Fasten the POWERPHOS to the C-arm. Make sure that the screws are tightened evenly.