Disassembly 2 - 9
Siemens AG SPR2-240.841.01 Page 9 of 12 POWERMOBIL
Medical Solutions Rev. 01 05.03 CS PS 24
Cable guide in the horizontal carriage 2
• Remove the cover above the cable (1/Fig. 17). Also loosen the cable clamps (2/Fig. 17).
• Unscrew the two cable guides, see (1/Fig. 18) and (2/Fig. 18).
• Unscrew the cable guide (1/Fig. 19).
• Unscrew the cable guide (1/Fig. 20). After moving the horizontal carriage, the second
screw of the cable guide will be accessible.
Fig. 17 Fig. 18
Fig. 19 Fig. 20