Installation 3 - 3
Siemens AG SPR2-240.841.01 Page 3 of 10 POWERMOBIL
Medical Solutions Rev. 01 05.03 CS PS 24
Run and connect the protective ground wire and the cable (1/Fig. 6) to the new cable
module. These cables also lie with one turn around the angulation axis.
Fasten the two cables to the cable module using cable ties (Fig. 6).
The two thick cables for the cable module have a red mark running round them (1/Fig. 7).
Insert and secure these cables together with the other cables of the cable module in the
clamping point (1/Fig. 8).
To check the cable lengths rotate the cable module angulation from end position to end
position. The cable loop around the angulation axis must be sufficiently long, but with
only slight excess length.
Fig. 6 Fig. 7
Fig. 8
Run the cables in the horizontal carriage from the cable module
towards the guide, and attach them.
NOTE