4 - 2 Checks and Adjustments
• Keep the C-arm in this position.
• Turn the bell (1/Fig. 5) until the rollers (2/Fig. 5) are aligned with the marking on the bell
(3/Fig. 5). Fig. 6 shows the bell (1/Fig. 6), the marking (3/Fig. 6) and the shaft with the two
rollers (2/Fig. 6) once again in a removed state.
• In this position of the bell and rollers, push the pinion gear back into the original position.
Make sure that the teeth of the gear engage.
• Retighten the screw (1/Fig. 3).
Fig. 5 Fig. 6