POWERMOBIL SPR2-240.841.01 Page 2 of 6 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
4 - 2 Checks and Adjustments
Keep the C-arm in this position.
Turn the bell (1/Fig. 5) until the rollers (2/Fig. 5) are aligned with the marking on the bell
(3/Fig. 5). Fig. 6 shows the bell (1/Fig. 6), the marking (3/Fig. 6) and the shaft with the two
rollers (2/Fig. 6) once again in a removed state.
In this position of the bell and rollers, push the pinion gear back into the original position.
Make sure that the teeth of the gear engage.
Retighten the screw (1/Fig. 3).