POWERMOBIL SPR2-240.841.01 Page 6 of 6 Siemens AG
Rev. 01 05.03 CS PS 24 Medical Solutions
4 - 6 Checks and Adjustments
Braking forces / braking moments 4
Check the braking forces/braking moments according to the service instructions
(chapter 7, Braking forces / braking moments).
POWERPHOS settings 4
Set the POWERPHOS and collimator according to the system service instructions
(chapter 6, "POWERPHOS").
Afterwards, refit the POWERPHOS cover.
Covers 4
Refit all covers. Make sure the protective ground wires make good contact.
Image quality 4
Perform the IQ quick test.
Ground Wire Test 4
Perform the protective ground wire test according to ARTD-002.731.17.
The protective round wire resistance must not exceed 0.2 ohms.