Service and maintenance
12.2 Error messages
SIMATIC RF185C, RF186C, RF188C, RF186CI, RF188CI
156 Operating Instructions, 04/2020, C79000-G8976-C512-03
4x 0xE1FE0400 Initialization error
Transponder is unable to execute the initialization command
Possible causes:
• Transponder is defective.
The transponder memory is full.
4x 0xE1FE0600 Error in transponder memory
The transponder has never been written to or has lost the contents of its memory due to battery
failure.
Possible causes / action to be taken:
• Replace the transponder (if the battery bit is set).
• Re-initialize the transponder.
4x 0xE1FE0800 The transponder in the antenna field does not have the expected UID / EPC ID or has no UID /
The command is not supported by the transponder.
The transponder is read/write-protected.
The transponder is not responding.
The transponder password is incorrect. Access is denied.
4x 0xE1FE8300 The verification of the written transponder data has failed.
Possible causes:
• Transponder is defective.
• Transponder is in the limit area.
General transponder error
4x 0xE1FE8500 The transponder has too little power to execute the command.
Possible causes:
• Transponder is in the limit area.