WARNING
Use in hazardous area
Risk of explosion.
• Only use equipment that is approved for use in the intended hazardous area and labeled
accordingly.
• Do not use devices that have been operated outside the conditions specied for hazardous
areas. If you have used the device outside the conditions for hazardous areas, make all Ex
markings unrecognizable on the nameplate.
WARNING
Use outside the permitted voltage limits
Explosion hazard
• Only use the device within the voltage limits specied on the nameplate.
WARNING
Formation of explosive sample gas mixtures
Explosion hazard
Air in the gas path can cause an explosion during the introduction of ammable gas mixtures
into the OXYMAT7.
• Prior to the introduction of ammable sample gas mixtures, ush the gas path in the
analyzer and the gas line with inert gas when ammable gas mixtures are introduced, as
follows:
– into the rack unit for any pressure
– into the wall-mounted device at a pressure above 100 hPa (rel.)
WARNING
Maintenance during continued operation in a hazardous area
There is a risk of explosion when carrying out repairs and maintenance on the device in a
hazardous area.
• Isolate the device from power.
- or -
• Ensure that the atmosphere is explosion-free (hot work permit).
Safety notes
2.5Use in hazardous areas
SIPROCESS GA700 rack and wall mounted devices
22 Compact Operating Instructions, 06/2023, A5E35134047-AB