WARNING
Wetted parts unsuitable for the process media
Risk of injury or damage to device.
Hot, toxic and corrosive media could be released if the wetted parts are unsuitable for the
process medium.
• Ensure that the material of the device parts wetted by the process medium is suitable for the
medium. Refer to the information in Technical data (Page87).
WARNING
Exceeded maximum permissible operating pressure
Risk of injury or poisoning.
The maximum permissible operating pressure depends on the device version, pressure limit
and temperature rating. The device can be damaged if the operating pressure is exceeded. Hot,
toxic and corrosive process media could be released.
Ensure that maximum permissible operating pressure of the device is not exceeded. Refer to
the information on the nameplate and/or in Technical data (Page87).
WARNING
Insucient air supply
The device may overheat if there is an insucient supply of air.
• Install the device so that there is sucient air supply in the room.
• Observe the maximum permissible ambient temperature. Refer to the information in the
section Technical data (Page87).
NOTICE
Incorrect mounting
The device can be damaged, destroyed, or its functionality impaired through improper
mounting.
• Before installing ensure there is no visible damage to the device.
• Make sure that process connectors are clean, and suitable gaskets and glands are used.
• Mount the device using suitable tools. Refer to the information in Technical data (Page 87).
NOTICE
Aggressive atmospheres
Damage to device through penetration of aggressive vapors.
• Ensure that the device is suitable for the application.
Installing/mounting
3.1Basic safety notes
SIPROCESS GA700 rack and wall mounted devices
26 Compact Operating Instructions, 06/2023, A5E35134047-AB