Test setup for leakage test of analyzer modules in all versions
To check for leaks, we recommend the test setup shown below:
;;
Shut-o󹪜 valve
Pressure regulator
Manometer (gauge pressure)
1 Sample gas inlet
2 Sample gas outlet
3 OXYMAT 7 / ULTRAMAT 7: Unscrewing blanking plugs
4 OXYMAT 7 / ULTRAMAT 7: Unscrewing blanking plugs
Figure5-2 Recommended test setup to check for leaks
Commissioning
5.4Checking gas paths for leaks
SIPROCESS GA700 rack and wall mounted devices
70 Compact Operating Instructions, 06/2023, A5E35134047-AB