3. Wait for transition to warming-up phase:
When the splash screen disappears, the device is in the warming-up phase. The current
warm-up progress is indicated by a bar chart in the dialog window.
4. Call up the measured value display during the warm-up phase:
Close the warming-up window by pressing <ENTER>.
The display shows the measured-value display in read mode.
For the duration of the warming-up phase, the device is in "Function check" mode. The
following symbol is displayed in the status bar:
Setting up
Note
Operator control options
When commissioning for the 󹪝rst time, two methods can be used. You can use the menu
"[1]Quick Start" to start the device with the simpli󹪝ed process or opt for the more detailed
procedure.
For information on the simpli󹪝ed procedure "Quick Start" refer to the operating instructions.
Make the necessary settings in the appropriate submenus.
Commissioning
5.5Local commissioning
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB 79