Gas inlet conditions
OXYMAT 7 normal temperature 0…60°C (32 ... 140 °F) with relative humidity <90%, avoid con‐
densation in the gas path (moisture content results in measurement
errors)
OXYMAT 7 high-temperature 0 to 130 °C (32 ... 266 °F), avoid condensation in the gas path (mois‐
ture content results in measurement errors)
1)
Optimum measuring results can only be achieved at 2000 hPa ±150 hPa above sample gas pressure. Calibrations are required
for any larger deviations.
Note
Composition of the reference gas
Using an unsuitable reference gas can cause the reference gas throttle (capillary tube) to be
closed. The device is then no longer capable of measuring.
Therefore, use dust-free reference gas or a standard dust 󹪝lter with a mean pore width of <5 µm
in the reference gas inlet.
Note
If 󹪞ammable gas mixtures are supplied in the rack mounted device or at a pressure > 100 hPa in
the wall mounted device, the gas line inside the OXYMAT analyzer and the supply line shall be
purged with inert gas prior to the analysis.
Note
The analysis of gas mixtures above UFL is only permitted with purging and at a pressure < 100
hPa in the wall device for OXYMAT module.
Table 7-12 OXYMAT 7: Temperature of sample chamber and gas path
Temperature of sample chamber
and gas path
OXYMAT 7 OXYMAT 7 high-temperature
1)
Gas path 5K above ambient temperature 72 ... 130°C (162 ... 266 °F), adjustable
Sample chamber 72°C (162 °F) 72 ... 130°C (162 ... 266 °F), adjustable
1)
The high-temperature version of the analyzer module is not intended for installation in the rack unit.
Table 7-13 OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport -30…70°C (-22 ... 158 °F)
Max. permissible ambient temperature surrounding the basic
device during operation
1)
0…50°C (32 ... 122 °F)
Ambient humidity (rel. humidity) during storage, transport or
operation
2)
<90%
1)
Restriction for installing an ULTRAMAT7 analyzer module (combined operation): 5…45°C (41 ... 113 °F)
2)
Avoid condensation on the installed components!
Technical data
7.1General technical speci󹪝cations
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB 93