24 General Information
SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG
07.05 CS PS 24
Page 24 of 46
Medical Solutions
3.2.1 Leakage current measurement 3. Electrical safety, check:
SIE Leakage current
5.1 Final work steps, check:
SIE Leakage current
3.2.2 Equivalent leakage current measure-
ment
n.a.
3.3 Patient leakage current or equivalent
patient leakage current measurement
n.a.
3.3.1 Patient leakage current measurement n.a.
3.3.2 Equivalent patient leakage current
measurement
n.a.
4. Mechanical checks n.a.
4.1 Are all wall, ceiling and floor mountings
secure and undamaged?
3. Mechanical safety, check:
SIM Monitor(s) / displays
4.2 Are all mechanically moved system
parts clean and running smoothly (lubri-
cated, if required)?
3. Mechanical safety, checks:
SIM Covers
SIM C-arm
SIM Lifting column
4.3 Are the wire cables, belts and spindles
free of wear and tear?
3. Mechanical safety, checks:
SIM Lifting column
SIM 3D reconstruction
4.4 Mobile equipment: Are the wheels, roll-
ers and brakes OK?
3. Mechanical safety, checks:
SIM Foot brake
SIM Brakes SIM Wheels
4.5 Is there any unusual noise during oper-
ation (e.g., gearing)?
3. Mechanical safety, checks:
SIM C-arm
SIM Lifting column
5. Functional checks n.a.
5.1 Function of the Emergency Stop switch 3. Mechanical safety, check:
Emergency Stop switch
3. Electrical safety, check:
SIE 3D reconstruction
Technical safety checks, test certificate
of the check
Maintenance instructions for the
check/comments