26 General Information
SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG
07.05 CS PS 24
Page 26 of 46
Medical Solutions
6. Product-specific checks 3. Electrical safety, checks:
SIE 3D reconstruction
SIE 3D reconstruction in combination with
a navigation system
SIE IQ quick test
SIM DICOM bridge IQ test
SIM Codonics printer IQ test
7. Test result/evaluation: Evaluate the long-term trends of the pro-
tective conductor resistance and the leak-
age current by comparing the current
measuring values to those of the preceding
technical safety checks. A sudden or unex-
pected increase in the measured values
may indicate a safety-related defect - even
if the limit values are not exceeded. Follow
the instructions for checks in the section on
final work steps.
Enter the results of the check.
Have the operator, or a person authorized
by the operator, sign the TSC report.
Hand the TSC report over to the operator,
or a person authorized by the operator.
Technical safety checks, test certificate
of the check
Maintenance instructions for the
check/comments