EasyManua.ls Logo

Siemens SITRANS F FST030 - Page 105

Siemens SITRANS F FST030
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
SITRANS F Ultrasonic flowmeters FST030 (HART)
Operating Instructions, 08/2017, A5E35971532-AC
103
ID/
Icons
Diagnostic
Action
Comment
212
CH4 Pulse overflow. Pulse output insufficient pulse
separation. Reduce pulses per amount, or
reduce pulse width, or increase amount
If Operation mode is config-
ured to Pulse output
214
CH2 Channel simulated. Disable 'Simulation' before
returning to normal operation
215
CH3 Channel simulated. Disable 'Simulation' before
returning to normal operation
216
CH4 Channel simulated. Disable 'Simulation' before
returning to normal operation
217
Process values frozen All process values and totalizers are frozen.
To return to normal operation the related input
signal has to switch.
218
Outputs forced Outputs are forced. To return to normal opera-
tion the related input signal has to switch
219
CH2 Loop current deviation. Check current output
cable connection. In passive mode check
power supply.
If Operation mode is config-
ured to Current output
220
CH3 Loop current deviation. Check current output
cable connection. In passive mode check
power supply.
If Operation mode is config-
ured to Current output
221
CH4 Loop current deviation. Check current output
cable connection. In passive mode check
power supply.
If Operation mode is config-
ured to Current output

Table of Contents

Related product manuals