Commissioning
6.3 Menu structure
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002 41
Lock field symbols
RESET MODE: Zero setting of totalizers and
initialization of setting
Keypad
The keypad is used to set the flowmeter. The keys function as follows:
This key (when held for 2 sec.) is used to switch between
operator menu and setup menu. In transmitter setup menu, a
short press will cause a return to previous level.
This key is used to step forward through the menus. It is the
only key normally used by the operator.
This key is used to step backwards through the menus.
With this key settings or numerical values are changed.
With this key figures to be changed are selected.
This key enables the operator to change settings and it gives
access to submenus.
6.3 Menu structure
The menu is built up of two parts. An operator menu and a setup menu, see also overview
diagrams MAG 5000/6000 and MAG 5000/6000 CT.
Operator menu
The operator menu is for daily operation. It is customized in the operator menu setup. The
transmitter always starts up in operator menu No. 1. The forward and the backward keys
are used to step through the operator menus.
Setup menu
The setup menu is for commissioning and service only. Access to the setup menu is gained
by pressing the top up key for 2 seconds. The setup menu operates in two modes:
● View mode
● Setup mode
View mode is a read-only mode. The pre-selected settings can only be scanned.