Commissioning
6.4 Changing password
SITRANS F M MAG 5000/6000
42 Operating Instructions, 12/2013, A5E02338368-002
Setup mode is a read and write mode. The pre-selected settings can be scanned and
changed. Access to the setup mode is password-protected. The factory set password is
1000.
Access to a submenu in the setup menu is gained by pressing the lock key . Press the top
up key briefly to return to the previous menu. Press longer (2 sec.) to exit the setup menu
and return to operator menu No. 1.
6.4 Changing password
The setup menu is password-protected in order to ensure that only authorized personnel can
make any changes in transmitter settings.
Change password as follows:
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key or backward key to reach password menu.
4. Press lock/unlock key to unlock password.
5. Use select key and change key to change password.
6. Press lock/unlock key to confirm new password.
7. Press top up key two times to exit setup mode.
See change password diagram.
The factory-set password is 1000, but it can be changed to any value between 1000 and
9999.
Factory setting of password can be re-established as follows:
1. Switch off power supply.
2. While pressing top up key - switch on power supply.
3. Release top up key after 10 sec.